| |
Gases
Always Follow the Path of Least Resistance
To test this dead
space, we conducted smoke tests on some of the more common masks and
nosecones we found in use in facilities. Smoke was introduced into
the fresh gas flow of a typical non-rebreathing tee circuit at a flow
rate of 1 liter per minute (LPM). The fresh gas flowed through the
fresh gas tube to the nosecone/endotracheal tube connection, then
continued out through the waste gas tube without saturating the patient's
nose area.
Dead
Space and Dead Patients
Meanwhile, the
animal continues to breathe. The static volume of air within the mask
(dead space) is sufficient to temporarily fill the animal's minute
volume needs. Unfortunately for the animal, the same air is being
inhaled and exhaled repetitively. In a short amount of time, the oxygen
in the dead space is depleted and replaced with CO2,
at which point the animal begins to breathe faster. A normal response
to this symptom would often be to turn up the vaporizer dial because
we assume that the patient is waking up. We probably wouldn't overdose
the patient since it isn't getting much fresh gas anyway, but we also
won't meet its metabolic requirements for oxygen and the patient could
die as a result.
A Solution
For most uses,
we recommend "bubble tubing" nosecones for rats, mice, gerbils
and hamsters. |
- They are
less likely to leak than other types.
- They have
no dead space.
- They allow
far more cranial access than the larger facemasks.
- They can
be stored in a cold sterile solution.
- You can make
your own - about 100 for less than $50! If you have an outbreak
and still need to do procedures, the ability to change nosecones
with each cage or with each animal can be invaluable.
|
|
Basically, a
nosecone must seal tightly around the patient's nose to avoid leaks
and it must be small enough to eliminate dead space. If you are
using masks with large amounts of dead space, sufficient fresh gas
delivery can be accomplished by increasing the oxygen flow rate.
However, it must be mentioned that increased flows may lead to increased
trace gas exposure.
|
|